| looking for Racal Dana 9087 service manual ? | ||
| Author yvling Views 84 Posted at 2008/8/29 15:45 [View In Forum] | ||
|
Hi, ufuwysfwb ufuwysfwb ufuwysfwb does anyone know how to get service/operator manual for ufuwysfwb ufuwysfwb ufuwysfwb The Racal Dana 9087 Rf signal Generatorufuwysfwb ufuwysfwb ufuwysfwb ufuwysfwb ufuwysfwb ufuwysfwb I recently bougth this generator and only "error 80" appear on the display.ufuwysfwb ufuwysfwb ufuwysfwb any help will be very appreciate.ufuwysfwb ufuwysfwb ufuwysfwb Thanks in advance |
| Reply No. | Replier | Total Replies 29 [View All Replies] [New Reply] | Replied at |
| 1 | ybyygu | You bought it on ebay , am I right ??? bloody heating problem ... I am looking for the same stuff patrick |
2008/8/29 15:45 |
| 2 | netlgc | Hi, Error code 80=Reference-generator loop out of lock. If you experience an error code 80, check rear internal/external frequency standard reference switch. Several have been found intermittant. OR internal oscilator 9442-5 problem. Check for 5 or 10mhz output on internal oscilator 9442-5. Takis |
2008/8/29 15:45 |
| 3 | chdd | I contacted RACAL EADS US and they told me the following about this rig : discontinued in 1998 made by RACAL UK which shutdown and nobody at RACAL's can service these rigs I gess the necessary fixture to reprogram the ROMS in case of loss have been scrapped Patrick |
2008/8/29 15:45 |
| 4 | mxg330 | F6iib wrote: Enquiry: looking for Racal Dana 9087 service manual ? can you send the pdfs of the service manuals regards Sorry Patrick I haven抰 got any PDF, only a photocopy from the instruction/technical manual. It is big and low quality for scanning, and it doesn抰 contain any information for the factory sealed modules (int.freq.standard, attenuator assembly, comb loop module, ref and LF synth. Module, output system module and FM system module). Maybe I can help you if you describe me the problem. Also years ago I made copies from the Processor board 10 PROMS, tell me if you need those files. My software revision number is 22.8 553-0 (SPECIAL FUNCTION 22). Regards Takis |
2008/8/29 15:45 |
| 5 | ai8957382 | that's ok for the PROM files .... can send to f6iib(at)free.fr regards Patrick |
2008/8/29 15:45 |
| 6 | dongyea | Done. | 2008/8/29 15:45 |
| 7 | yvling | HI thanks your all for your very appreciate help. i've bought the maintenence manual from a uk seller at the modest price of 60 Gpb. when it arrive i'll check my generator |
2008/8/29 15:45 |
| 8 | z315 | I didn't buy a RACAL 9087 A ebay member told me the Following : "I know Racal no longer support these unit and have given all test equipment to Electroservices Ltd Uk. Repair of these units is between ?00 and ?500. Over the next 18 months the MOD as about '500' of these units to get ride off !!!....some inside information ..." I Guess the MOD is the "Ministry of Defense" so there should be a lot of these units around ....if this guy says true stuff The guy , papko, told me he had a palette for sale .... I you happen to have the time to scan the Manual send me a copy just in case one day I buy one .... If I were I would scan that stuff and sell the CD on ebay .... lol to get some little money back I already have RACAL 1992 Freq counter and all the service/user doc in pdf If I can buy one ( 9087 ) for 30 quids , less then the carriage fees , could be worth if there is no errors whatever .... Hope you can enjoy using this nice equipment Patrick Added after 1 minutes: |
2008/8/29 15:45 |
| 9 | fujun2013 | someone who used to service these rigs gave me the following information I quote "error 80 ... there are 2 fault conditions with error 80 attached ,he give me clue as the fault and it would just the cost of a switch this can be verified by a 10mhz 1vpep sig to rear input port this should clear the fault 80.if still faulty then it's ref module there ae 2 osc in the unit that causes the lock " hope helps ... did u get the doc ??? |
2008/8/29 15:45 |
| 10 | spacebac | hi ya everyone. Thank your all for your help, i hope this topics will be usefull also for other unlucky ppls who bought the generator with this error Sorry Patrick i didn't received the manual yet,I hope to have it next week, I let you know. Anyway i'm starting to try to fix the generator. 10 Mhz @ 1Vpp present 5 Mhz From Xtal oscillator present. Injected 10 Mhz 1- 1,5Vpp to Ext ref :cry:Still same problem " Error 80!" Any suggestion |
2008/8/29 15:45 |